ILIM-7 IntelliLogger Analog Input Channel Expansion Module

Isolated 8 channel analog input expansion module for the IntelliLogger™ system.

ILIM-7 Analog Input Module
ILIM-7 Key Features
Adds 8 High Resolution Isolated analog inputs to IntelliLogger
All channels completely Kelvin isolated simplifying signal and transducer Connectivity
User programmable for uVdc, mVdc, mAdc and thermocouple direct input
Integral cold junction compensation for thermocouple readings
Integral over-current protection
Simple BBus cable connection for remoting modules away from the ILL base
Automatically recognized by HyperWare-II Software for intuitive setup

Product Images

Dip switch for BBus Module

Dip switch for BBus Module

ILIM-7 BBus Connection Ports

IL-80 shown with two expansion modules attached

IL-80 shown with two expansion modules attached

IL-20 with ILIM-7 modules daisy chained via BBus

IL-20 with ILIM-7 modules daisy chained via BBus

Product Information

Accessories

BPB-1 BBus Power Booster

BPB-1 BBus Power Booster

Injects supply voltage in BBus network for applications with 5 or more Expansion Modules or long BBus cable runs.

Expansion module attached to the IntelliLogger base with angle attachment bracket

Module Mounting “Stair Step” Bracket

Allows stacking of Expansion Modules and/or attachment to the ILL base.

Surface mounting plate. Bracket mounts to back of Data Logger and extends above and below.

Surface Mounting Bracket

For surface mounting of an Expansion Modules to a wall or enclosure back.

Analog Power Input Protection Fuses

50mA Input Protection Fuses

Fuses protect the analog inputs from over-current when configured as mAdc inputs. Replacement fuses are available in 5 packs.

BBus Termination Stub

Bbus Termination Stub:

The last ILIM-x module in a BBus daisy-chain should have a BBus Termination Resistor stub installed. This stub serves to terminate the RS-485 serial communication link which mimimizes communication errors with the ILL System Base. One Stub is provided at order time.