Overview
The HLP-10 is a self-contained portable power logging instrument
and energy monitoring system. It is the ideal instrument for energy
efficiency studies because it can include energy and the results
of that consumption for total system analysis. The HLP-10 can be
deployed to monitor the power consumption from compressors, chillers,
and any AC motors as well as the pressure, flow and temperatures
resulting from the energy expended for a more complete efficiency
study. In-plant efficiencies and sizing analysis can now be more
thoroughly performed by inclusion of all system components and parameters.
The Logic Beach HyperLogger PowerPlus logs AC voltage, current
and power logging capabilities and allows the addition of standard
industrial sensors to record temperature, flow, or pressure simultaneously.
The power data is transmitted from the Power Sensing Harness to
the data logger via an isolated serial data link where it is stored
along with other user-connected sensors or signal inputs. The HLP-10
is capable of monitoring power from 16 unique systems simultaneously
allowing for faster analysis and results.
The Power Sensing Harness split core transducers contain the electronics
to monitor and convert the measured parameters to the serial data
stream. An integral serial cable then connects to the HyperLogger
data logging system, which can be located a distance from the power
harness and it's associated high-voltage connections. The HLP-10
is programmed from within the Logic Beach HyperWare software application.
HyperWare is the revolutionary, market proven icon-based programming,
communication, real-time trending and data analysis software used
with all of Logic Beach's portable and remote site data logging
and alarming systems.
The HLP-10 is also available with a 500mA rechargeable power supply
for regulated sensor excitation power, model HLP-200. There are
two supplies each 250mA and capable of 3.5vDC to 22vDC output. For
details on the power supply see the RPS-1 data sheet. The HLP-200
incorporates this power supply with the HLP-10 similar to the HL-200
and HL-1 system.
|
Applications
- Compressor efficiency & sizing studies
-
Cogeneration
performance analysis
-
Chiller
efficiency studies
-
Time
of day power usage profiling
-
Power
Factor correction programs
-
Refrigeration
efficiency evaluations
- Voltage sag/surge detection
-
Transformer
Efficiency/Performance Testing
|
Features
-
Two and three phase operation
-
Monitor
up to 16 power feeds
-
Isolates
high voltage connections
-
Measure
up to 500VAC direct
-
Select
from 18 different clamp sizes
-
Up
to 2400 Amps AC
-
Local
readout
-
1%
measurement accuracy class
-
Icon based setup
-
Modem
/ RF /Cellular Communications
-
Programmable
Alarms
|
|
Measured Power Parameters
-
kWH: Consumption
- kW:
Demand
-
kW:
demand each phase*
-
Demand:
Avg., Min., Max.
-
Voltage,
line to line
-
Voltage
phase to phase
- Voltage:
line & phase to neutral*
-
Power factor
-
Power factor each phase*
-
Amps:
avg., current per phase
-
kVAR:
reactive power
-
kva:
apparent power
-
*Based
upon derived neutral voltage
|
Analog Signals
(when
used with optional interface Modules)
- Voltage
DC: +/- 30Vdc
- Current
DC: 0 - 22mA
- Temperatures,
T'couples, RTDs
- Frequency, Events, Pulses,
Resistance
|
| Only want to monitor current,
kilowatts, or kilowatt hours? Select just the right sensors for
your application here.
|
Operation
The power sensing harness is a self-contained
system that is comprised of the current clamps, fused voltage leads
and an electronics module. All the necessary circuitry is contain within
these components for processing of the voltage and current signals and
converting the voltage and current measurements into over 26 different
energy consumption, power, voltage and phase information. This information
is output, in serial format, to the data logger where the data logging
program reads and records the desired measurements into logger memory.
The HLP-10 HyperLogger is simply programmed from a PC running
the Logic Beach HyperWare data logger software. With the ease of HyperWare's
icon-based drag and drop programming method, icons representing phase
voltages, currents, power, etc are easily interconnected with a mouse
to build a data logging and alarming strategy. During operation, electrical
data from the power module as well as other connected sensor/signal
inputs (e.g. thermocouples, RTD's, thermistors, pressure, flow, 4-20mA,
etc) are logged to the HyperLogger's memory for later download
and analysis.)
Optional Interface Modules
| HLIM-1:
Analog Interface Module |
Thermocouples:
J,K,E,T,R,S
DC Voltage: 11 ranges, +/- 20mV to +/- 30V
DC Current: 7 ranges, +/- 200uA to +/- 22mA
|
| HLIM-2:
Digital Interface Module |
Frequency: 5Hz to
30kHz (square or sine wave)
Event: Contact closure or TTL, 0-15Vdc
Count: Contact closure or TTL to 20kHz |
| HLIM-4:
Resistance Interface Module |
RTD: 100 and 1000
Ohms
Thermistors: 10k Ohm @ 25degres C
Resistance: 12 ranges, 200ohms to 400K full scale |
| HLIM-5:
Memory/Modem Interface |
Module has internal
socket for removable memory cards from 60,000 to 600,000 samples
with capabilities for internal 14.4 modem. |
| HLIM-7:
Isolated Interface Module |
Thermocouples: J,K,E,T,R,S
DC Voltage: 6 ranges, +/- 23mV to +/- 30V
DC Current: 5 ranges, +/- 200uA to +/- 22mA |
| HLIM-8:
Eight Channel Digital Interface |
Eight individually
user-selectable channels configured as either inputs or outputs. |
| HLIM-9 |
Plug-in module for serial-bus
data recording. J1587 CAN bus, variations on Modbus and Mettler
Toledo protocols now available. Contact factory for others. |
Options
| Removable
Memory Cards |
PCMCIA memory cards allows for
easy removal and transport of collected data. Additionally, cards
are available to expand memory capacity to in excess of 500,000
samples. Cards require that HyperLogger have an HLIM-5 Interface
Module installed. |
| PSM-2;
Sensor/Loop Power Supply |
For applications requiring sensor
or loop power/excitation greater than the HyperLogger's internal
5Vdc source, the door-mounted PSM-2 can provide 10, 15, or 24Vdc
excitation voltages stepped up from the HyperLogger system
batteries. |
| HLP-200 |
An integrated system that consists
of a HyperLogger™ PowerPlus™ System Base and an RPS-1
rechargeable power supply integrated into a single NEMA 4X rated
enclosure. |
| Cell
Phone Option |
The HyperLogger can be supplied
with a cell phone option... resulting in a truly autonomous data
logging platform. |
|
Ethernet
Communication
|
Connect the HyperLogger to a LAN. |
| HAP;
Polling Sofware |
HAP is an accessory software application
that automates the communication, data downloading, file conversion
and even report generation process from one or multiple HyperLogger
and/or ModuLogger sites. |
Technical Specifications
DATA LOGGER
| Data Storage
Memory |
Battery-backed up
SRAM. Apx. 40,000 samples internal, to 600,000 samples with optional
PCMCIA removable memory card |
| Data Memory
Backup |
Lithium cell, 1 year
@ 25C |
| Memory Utilization |
User programmable;
Stop when Full or Rotary (FIFO) memory |
| A/D Converter |
12 bit plus sign (13
bit) SAR converter, 18 bit for HLIM-7 Interface Module |
| A/D Converter
Accuracy |
+/- 0.1% RDG + 1 bit |
| Sampling Throughput
Rate |
150+ Samples/Sec.,
rate varies with number and type of channels and programmed signal
processing |
| Interrface
Module Ports |
6 ports for plug-in
Interface Modules and special function modules (modem, PCMCIA) |
| Digital Port |
Integral General Purpose
Digital Input channel. Programmable for Event or high-speed Counter.
Contact closure or driven input |
| Outputs |
2 low-voltage N/O
relays; 500mA rated: S/W controlled 3 Digital (0/5 Vdc), Current
Ltd., Software controlled 5Vdc regulated output, 125 mA current
limited |
| Display |
Two line, 16 character
per line LCD |
| Clock |
Date and Time, 24
hour, battery backed up. |
| Glitch Recover |
Hardware watchdog
reset followed by software restart of last operation. |
| Power Consumption |
9 VDC nominal. Apx.
3mA between readings; apx. 50mA during readings; provided by 6
internal D-cells. |
| External Power |
9-16 VDC, 10-20 VAC
from any semi-regulated external source. Transzorb protected. |
POWER SENSING HARNESS
| Input primary
voltage |
208 to 480 VAC rms
Number of phases monitored: one or three |
| Frequency |
50/60 Hz |
| Primary current |
up to 2400A continuous/phase |
| Internal isolation |
2000 VAC rms |
| Case insulation |
600 VAC rms |
| Temperature
range |
0 - 60C° |
| Humidity range |
0 - 95% non-condensing |
| Accuracy |
1% |
| Output |
RS 485, 2-wire shield |
| Current transformers |
split core: 100, 300, 400, 800,
1600, 2400 |
| Over Current |
10% of rated current continuous |
| Inrush
Current |
5X rated current for
2 seconds |
|
|
|

Data